TDOCT with Optical Delay Line based on Static Helicoid Mirror


Por: Kolokoltsev O., Alejo M.A., Arista I.G., Qureshi N., Mejia E.

Publicada: 1 ene 2025
Resumen:
© 2025, Universidad Nacional Autonoma de Mexico. All rights reserved.In this work, we described a design of Optical Delay Line (ODL) for Time Domain Optical Coherence Tomography (TDOCT). ODL is based on a static helicoid mirror, integrated with an optical deflector. The delay line can provide up to 10 mm scanning depth range, thanks to its multi-pass configuration. The TDOCT system has been analyzed and optimized to operate in ultrahigh depth resolution regime. The estimated sensitivity of the TDOCT, at the presence of the relatively high level of the photon excess noise, can be of the order of-94 dB.

Filiaciones:
Kolokoltsev O.:
 Instituto de Ciencias Aplicadas y Tecnología, Universidad Nacional Autónoma de México, CU 04510, Cd. México, Mexico

Alejo M.A.:
 Instituto de Ciencias Aplicadas y Tecnología, Universidad Nacional Autónoma de México, CU 04510, Cd. México, Mexico

Arista I.G.:
 CONAHCYT-INAOE, Luis Enrique Erro # 1, Tonantzintla, Puebla, Mexico C.P. 72840

Qureshi N.:
 Instituto de Ciencias Aplicadas y Tecnología, Universidad Nacional Autónoma de México, CU 04510, Cd. México, Mexico

Mejia E.:
 Instituto de Ciencias Aplicadas y Tecnología, Universidad Nacional Autónoma de México, CU 04510, Cd. México, Mexico
ISSN: 16656423
Editorial
UNIV NACIONAL AUTONOMA MEXICO, APARTADO POSTAL 70-399, MEXICO CITY, D.F. 21, MEXICO, México
Tipo de documento: Article
Volumen: 23 Número: 6
Páginas: 638-650

MÉTRICAS