B1 uniformity correction of a slotted end-ring coil with mean-shift filtering at 4 Tesla


Por: Ramon Jimenez, Juan, Solis-Najera, Sergio, Tomasi, Dardo, Odon Rodriguez, Alfredo

Publicada: 1 ene 2016
Categoría: Physics and Astronomy (miscellaneous)

Resumen:
The quality of magnetic resonance images is greatly affected by the field uniformity of the RF coil. This problem used to detect the signal. There is a growing interest in high field magnetic resonance imaging mainly in applications involving animal models to study a variety of important diseases, where the image quality is of paramount importance. In the past 20 years a number of digital-processing image methodologies have been proposed to overcome this problem. Among these techniques, the mean-shift scheme has proved to be an efficient and reliable method at intermediate field magnetic resonance imaging. The goal in this work is to apply the mean-shift scheme to phantom images with poor uniformity acquired with whole-body magnetic resonance imager at 4 Tesla. © 2016 Author(s).

Filiaciones:
Ramon Jimenez, Juan:
 Univ Autonoma Metropolitana Iztapalapa, Dept Ingn Elect, Cd Mx 09340, Mexico

Solis-Najera, Sergio:
 Univ Nacl Autonoma Mexico, Fac Ciencias, Cd Mx 04510, Mexico

Tomasi, Dardo:
 Brookhaven Natl Lab, Dept Med, Upton, NY 11973 USA

Odon Rodriguez, Alfredo:
 Univ Autonoma Metropolitana Iztapalapa, Dept Ingn Elect, Cd Mx 09340, Mexico
ISSN: 0094243X
Editorial
AMER INST PHYSICS, 2 HUNTINGTON QUADRANGLE, STE 1NO1, MELVILLE, NY 11747-4501 USA, Estados Unidos America
Tipo de documento: Conference Paper
Volumen: 1747 Número:
Páginas:
WOS Id: 000380810000033